Detailed introduction of thermal shock test chamber: specifications, technical parameters, etc.

Detailed introduction of thermal shock test chamber: specifications, technical parameters, etc.

The thermal shock test chamber is divided into two-box and three-box, the difference being the internal structure and test method of the equipment. The equipment is mainly used to test the material structure or composite material, and it can test the chemical change or physics caused by thermal expansion and contraction in the shortest time in the short-term continuous environment under extremely high temperature and extremely low temperature. hurt.

Cold and thermal shock test chamber - sincerely recommend to you [forest frequency] brand! The following is the details of the device, please refer to:
The thermal shock test chamber meets the standard: GB/T2423.1.2, GB/T10592-2008, GJB150.3 high and low temperature impact test. This test equipment is an essential test equipment for metal, plastic, rubber, electronics and other materials industries, used for testing. The degree to which a material structure or composite material can withstand in a transient environment of extremely high temperature and extremely low temperature.

In the shortest time, the chemical or physical damage caused by thermal expansion and contraction is detected.

I. Specifications and technical parameters Temperature impact two-box Model: LP/2XCJ-100A Studio size: 450 × 450 × 500 Dimensions: 1560 × 870 × 1545
Model: LP/2XCJ-225A Studio size: 600 × 600 × 600 Dimensions: 1710 × 1020 × 1845
Model: LP/2XCJ-500A Studio size: 800 × 800 × 800 Dimensions: 1910 × 1220 × 2265
Model: LP/2XCJ-010A Studio size: 1000 × 1000 × 1000 Dimensions: 2110 × 1420 × 2665

Temperature shock three-box model: LP/3XCJ-100A Studio size: 450 × 450 × 500 Dimensions: 1560 × 870 × 1545
Model: LP/3XCJ-225A Studio size: 600 × 600 × 600 Dimensions: 1710 × 1020 × 1845
Model: LP/3XCJ-500A Studio size: 800 × 800 × 800 Dimensions: 1910 × 1220 × 2265
Model: LP/3XCJ-010A Studio size: 1000 × 1000 × 1000 Dimensions: 2110 × 1420 × 2665

Second, technical indicators
1, temperature range: high temperature box: RT ~ 150 ° C (no load)
Cryogenic box: RT ~ -20 ° C (no load)
2. Temperature fluctuation: ±1°C
3, temperature error: ± 3 ° C
4, heating rate: from normal temperature ~ 150 ° C ≤ 25min (full range average)
5, cooling rate: from normal temperature ~ -20 ° C ≤ 15min (full range average)
6, low temperature limit temperature: ≤ -25 ° C;
7, impact temperature: 150 ° C ~ -20 ° C
8, load: ≤ 30kg
9, two-box sample basket effective size:
250×250×250mm (LP/2XCJ-100A)
400×400×450mm (LP/2XCJ-225A)
600×600×600mm (LP/2XCJ-500A)
800×800×750mm (LP/2XCJ-010A)
10, temperature recovery time: ≤ 5min
11, sample rack conversion time: ≤ 15s
12, power: 13KW/15KW/18KW/20KW

Third, the box structure
1. The outer wall material of the box: the outer surface steel plate is sprayed;
2, the body wall material: SUS304 stainless steel plate;
3. Observation window: Install a rectangular multi-layer hollow insulating glass hot glass observation window (300×200mm) in the high temperature box. Install the lighting on the observation window;
4. Insulation material: Polyurethane foam plastic is connected to the inner chamber, outer shell, high greenhouse and low greenhouse by high temperature and high pressure foaming;
5, sample rack bearing: no more than 20 kg;
6. Sample rack moving mode: The sample rack is moved up and down by means of a linear reciprocating drive system at the bottom of the mounting box;

Fourth, measurement and control system
1. Control device: The main controller adopts South Korea imported "TEMI8226" dual-loop high-precision-liquid crystal display-touch button temperature controller, programmable microcomputer PID control SSR output operation, large LCD color touch screen screen, simple screen operation, program Easy to edit;
2, accuracy: 0.1 ° C (display range);
3. Resolution: ±0.1 °C;
4. Refrigeration system and compressor: In order to ensure the cooling rate and minimum temperature of the test box, the test box adopts a binary cascade air-cooled refrigeration system consisting of imported German semi-closed compressors. The cascade refrigeration system comprises a high temperature refrigeration cycle and a low temperature refrigeration cycle. The connecting container is an evaporative condenser. The evaporating condenser is also used for energy transfer. The heat energy in the working chamber is transmitted through the two-stage refrigeration system to achieve the enthalpy temperature. the goal of. The design of the refrigeration system uses energy regulation technology, and an effective treatment method can ensure the energy consumption and cooling capacity of the refrigeration system can be effectively adjusted under the normal operation of the refrigeration unit, so that the operation cost of the refrigeration system And the failure rate drops to a more economical state;
5, air conditioning mode: forced ventilation internal balance temperature adjustment method (BTC). The method means that when the refrigeration system is continuously operated, the control system controls the output of the heater according to the result of the PID automatic calculation output according to the set temperature point, and finally achieves a dynamic balance.